ProductGuru Recalls

Official recall record

Histocore Pegasus (ref 75407)

Health Canada — Medical Devices · 2024-04-18 · CanadaType II
Issuing authority
Health Canada — Medical Devices
Published
2024-04-18
Country of alert
Canada
Alert level
Type II
Alert type
recall

What the authority reported

Leica Biosystems became aware, through post market surveillance, of an issue regarding poorly processed or damaged biopsy tissue specimens on the Histocore Pegasus / Histocore Pegasus Plus, resulting from reagent levels exceeding the maximum fill level marks on reagent bottles or in the paraffin tanks. This recall is to inform customers of this issue and actions to be taken. Recall start date: April 3, 2024

What to do

Follow the instructions in the official notice — recall remedies (refund, repair, replacement, or disposal) are defined by the issuing authority and the responsible business, and they differ per recall. Do not rely on summaries, including this one.

Verify at the source: Health Canada — Medical Devices — official record →
Record last checked against the registry on 2026-05-31.

More recalls: Health Canada — Medical Devices · Canada · all filters

ProductGuru Recalls republishes official safety information with attribution. We are not the issuing authority and this page is not advice. If this product is a medical device or medication, speak to a doctor or pharmacist. Found an error? Report it — corrections are published.