Airvo 2 and MyAirvo 2 Humidifier
- Issuing authority
- Health Canada — Medical Devices
- Published
- 2024-04-24
- Country of alert
- Canada
- Alert level
- Type II
- Alert type
- recall
What the authority reported
In devices manufactured before 14 August 2017, the speaker configuration may result in distorted, intermittent or inaudible alarm sound levels. This does not affect the therapy delivered by the Airvo 2 or MyAirvo 2 device and the devices will otherwise perform as intended. Recall Start Date: April 3, 2024