iMRI 3T S

Health Canada — Medical Devices · 2024-04-25 · CanadaType II
Issuing authority
Health Canada — Medical Devices
Published
2024-04-25
Country of alert
Canada
Alert level
Type II
Alert type
recall

What the authority reported

The recall is being initiated to address a notification received by Deerfield Imaging, Inc. from Siemens regarding potential dot- or line-shaped artifacts caused by the presence of lubricating grease within the headband of the over-ear headphones when used during head examinations with the Magnetom MRI systems. Siemens became aware of this issue during post market surveillance. Siemens has identified that the root cause of the issue was a manufacturing processing error. Recall Start date: April 12, 2024