Official recall record
iMRI 3T S
- Issuing authority
- Health Canada — Medical Devices
- Published
- 2024-04-25
- Country of alert
- Canada
- Alert level
- Type II
- Alert type
- recall
What the authority reported
The recall is being initiated to address a notification received by Deerfield Imaging, Inc. from Siemens regarding potential dot- or line-shaped artifacts caused by the presence of lubricating grease within the headband of the over-ear headphones when used during head examinations with the Magnetom MRI systems. Siemens became aware of this issue during post market surveillance. Siemens has identified that the root cause of the issue was a manufacturing processing error. Recall Start date: April 12, 2024