What the authority reported

It has been discovered that when the user copies the customized event categories from one unit to another, the event catalog displays the customized event categories, but the changes are not actually applied. If a user reviews the events in the event monitor the user will see in the event details that the event category is not updated to reflect the customization. This issue is the result of cached changes to the event catalog not propagating in the database. Recall Start Date: April 17, 2024