Official recall record
Patient Information Center iX (ref 75444)
- Issuing authority
- Health Canada — Medical Devices
- Published
- 2024-04-25
- Country of alert
- Canada
- Alert level
- Type II
- Alert type
- recall
What the authority reported
It has been discovered that when the user copies the customized event categories from one unit to another, the event catalog displays the customized event categories, but the changes are not actually applied. If a user reviews the events in the event monitor the user will see in the event details that the event category is not updated to reflect the customization. This issue is the result of cached changes to the event catalog not propagating in the database. Recall Start Date: April 17, 2024
What to do
Follow the instructions in the official notice — recall remedies (refund, repair, replacement, or disposal) are defined by the issuing authority and the responsible business, and they differ per recall. Do not rely on summaries, including this one.
Record last checked against the registry on 2026-05-31.