Official recall record
Philips Respironics BiPAP and Humidifiers
- Issuing authority
- Health Canada — Medical Devices
- Published
- 2024-05-02
- Country of alert
- Canada
- Alert level
- Type I
- Alert type
- recall
What the authority reported
The affected products feature a ventilator inoperative alarm, which occurs when the ventilator detects an internal error or a condition that may affect therapy. this may manifest in the following ways: The device may reboot intermittently for 5-10 seconds (stops providing therapy, screen goes blank during the reboot, and there is a single audible alert), restarting therapy, and returning to delivering therapy with same patient settings. or The device may reboot intermittently for 5-10 seconds (stops providing therapy, screen goes blank during the reboot, and there is a single audible alert), restarting therapy, and returning to delivering therapy but with factory default settings. or When there are three (3) reboots within a 24-hour period, the device will enter a ventilator inoperative state (therapy stopped, audible and visual alarms present). or The device may enter a ventilator inoperative state without a reboot preceding this condition. Any of the above scenarios could result in interruption and/or loss of therapy which may lead to hypoventilation, mild to severe hypoxemia, hypercarbia, respiratory failure/insufficiency, or potentially death in the most vulnerable patients. Recall start date: March 19, 2024
What to do
Follow the instructions in the official notice — recall remedies (refund, repair, replacement, or disposal) are defined by the issuing authority and the responsible business, and they differ per recall. Do not rely on summaries, including this one.
Record last checked against the registry on 2026-05-31.