EsoFLIP Dilation Catheter 30mm

Health Canada — Medical Devices · 2024-05-03 · CanadaType I
Issuing authority
Health Canada — Medical Devices
Published
2024-05-03
Country of alert
Canada
Alert level
Type I
Alert type
recall

What the authority reported

Medtronic has received complaints of ES-330 catheters measuring diameters that are significantly (more than 50%) lower than actual. The failure mode may be reported as "failed precheck", "incorrect distensibility reading", "no distensibility value displayed", "adverse event without identified device or use problem", or "diameter incorrect." Medtronic Lafayette returned product analysis lab testing of unused, returned ES-330 catheters showed saline conductivity that is approximately 10 times lower than specification. The ES-330 saline conductivity specification is 2.25 mS/cm. The saline conductivity of multiple unused, returned ES-330 catheters had saline conductivity at or near 0.25 mS/cm. Recall start date: April 25, 2024

What to do

Follow the instructions in the official notice — recall remedies (refund, repair, replacement, or disposal) are defined by the issuing authority and the responsible business, and they differ per recall. Do not rely on summaries, including this one.

Verify at the source: Health Canada — Medical Devices — official record →
Record last checked against the registry on 2026-05-31.

More recalls: Health Canada — Medical Devices · Canada · all filters

ProductGuru Recalls republishes official safety information with attribution. We are not the issuing authority and this page is not advice. If this product is a medical device or medication, speak to a doctor or pharmacist. Found an error? Report it — corrections are published.