Official recall record
IOMAX Cortical Module (ref 75489)
- Issuing authority
- Health Canada — Medical Devices
- Published
- 2024-05-07
- Country of alert
- Canada
- Alert level
- Type II
- Alert type
- recall
What the authority reported
An assembly error was discovered on the cortical module (PN 190296-200). A cortical module insulator that is used to provide an electric barrier between circuit boards for TCS-9, ICS, and other internal parts was found to be installed incorrectly. The insulator provides clearance between safety critical components and incorrect installation does not meet the minimum creepage/air gap required to meet IEC 60601-1. Recall start date: April 23, 2024
What to do
Follow the instructions in the official notice — recall remedies (refund, repair, replacement, or disposal) are defined by the issuing authority and the responsible business, and they differ per recall. Do not rely on summaries, including this one.