Official recall record
Presource® Kits Containing Curity® Flexible Adhesive Bandage X-large Components
- Issuing authority
- Health Canada — Medical Devices
- Published
- 2024-05-07
- Country of alert
- Canada
- Alert level
- Type II
- Alert type
- recall
What the authority reported
Cardinal Health recently initiated a recall on a specific lot of the Curity® Flexible Adhesive Bandage X-large (recall reference no 2024FCA21). The recall was issued due to potential contamination with latex adhesive residual. Recall start date: April 24, 2024
What to do
Follow the instructions in the official notice — recall remedies (refund, repair, replacement, or disposal) are defined by the issuing authority and the responsible business, and they differ per recall. Do not rely on summaries, including this one.
Verify at the source:
Health Canada — Medical Devices — official record →
Record last checked against the registry on 2026-05-31.
Record last checked against the registry on 2026-05-31.
ProductGuru Recalls republishes official safety information with attribution. We are not
the issuing authority and this page is not advice. If this product is a medical device or
medication, speak to a doctor or pharmacist. Found an error?
Report it — corrections are published.