ProductGuru Recalls

Official recall record

Presource® Kits Containing Curity® Flexible Adhesive Bandage X-large Components

Health Canada — Medical Devices · 2024-05-07 · CanadaType II
Issuing authority
Health Canada — Medical Devices
Published
2024-05-07
Country of alert
Canada
Alert level
Type II
Alert type
recall

What the authority reported

Cardinal Health recently initiated a recall on a specific lot of the Curity® Flexible Adhesive Bandage X-large (recall reference no 2024FCA21). The recall was issued due to potential contamination with latex adhesive residual. Recall start date: April 24, 2024

What to do

Follow the instructions in the official notice — recall remedies (refund, repair, replacement, or disposal) are defined by the issuing authority and the responsible business, and they differ per recall. Do not rely on summaries, including this one.

Verify at the source: Health Canada — Medical Devices — official record →
Record last checked against the registry on 2026-05-31.

More recalls: Health Canada — Medical Devices · Canada · all filters

ProductGuru Recalls republishes official safety information with attribution. We are not the issuing authority and this page is not advice. If this product is a medical device or medication, speak to a doctor or pharmacist. Found an error? Report it — corrections are published.