ProductGuru Recalls

Official recall record

Stryker ProCuity Beds

Health Canada — Medical Devices · 2024-05-08 · CanadaType II
Issuing authority
Health Canada — Medical Devices
Published
2024-05-08
Country of alert
Canada
Alert level
Type II
Alert type
recall

What the authority reported

Stryker has identified that the manufacturing records for certain ProCuity beds are missing test values for electrical safety tests as required by IEC 62353:2014. As a result of the missing test data, the impacted units are not compliant to the above standard. Therefore, Stryker is initiating this voluntary recall to perform the necessary testing and documentation in the field. Recall start date : April 29, 2024

What to do

Follow the instructions in the official notice — recall remedies (refund, repair, replacement, or disposal) are defined by the issuing authority and the responsible business, and they differ per recall. Do not rely on summaries, including this one.

Verify at the source: Health Canada — Medical Devices — official record →
Record last checked against the registry on 2026-05-31.

More recalls: Health Canada — Medical Devices · Canada · all filters