Official recall record
InstaClear Sheath
- Issuing authority
- Health Canada — Medical Devices
- Published
- 2024-05-17
- Country of alert
- Canada
- Alert level
- Type II
- Alert type
- recall
What the authority reported
Olympus (Gyrus Acmi, Inc.) is initiating a medical device recall action after investigating a complaint for an InstaClear Sheath (model # LCS4K30BTOL) in which the plate on the distal end of the sheath broke off during a procedure. Olympus has received sixty-seven (67) complaints across all sheath models related to breakage or damage to the pin or plate at the distal end of the InstaClear Sheath since the product was launched in 2016. Four (4) complaints were reported as serious injuries. Recall Start Date: May 3, 2024