Plum A+3 Infusion Pumps

Health Canada — Medical Devices · 2024-05-23 · CanadaType II
Issuing authority
Health Canada — Medical Devices
Published
2024-05-23
Country of alert
Canada
Alert level
Type II
Alert type
recall

What the authority reported

Expanded recall notice to include (Plum A+3 infusion pumps). If a Plum 360 or Plum A+ infusion system is running on battery power, a low battery and depleted battery alarm should typically activate when thirty minutes and three minutes, respectively, of estimated battery runtime remain. In addition, when the pump detects a loss of battery capacity, the pump will display a message to replace the battery. On a Plum 360 pump, the screen will display "keep plugged into ac! service battery / replace pump" and Plum A+ pump will display "warning: replace battery." Due to a manufacturing defect from the battery supplier, affected batteries may experience a loss of capacity earlier in the battery lifecycle than expected and overall battery runtime may decrease earlier in the battery lifecycle than expected. If an affected battery is not replaced when the pump displays the message to replace the battery, there may be less than thirty minutes of battery runtime after the low battery alarm resulting in the depleted battery alarm occurring sooner than expected. At this point, the pump stops delivery, alarms depleted battery for three minutes, and the pump turns off. Please note: this recall may also affect spare batteries from CSB, manufactured before January 1, 2023. If the first two characters on the label are 22 or lower, the battery is potentially affected. Recall start date: May 5, 2024

What to do

Follow the instructions in the official notice — recall remedies (refund, repair, replacement, or disposal) are defined by the issuing authority and the responsible business, and they differ per recall. Do not rely on summaries, including this one.

Verify at the source: Health Canada — Medical Devices — official record →
Record last checked against the registry on 2026-05-31.

More recalls: Health Canada — Medical Devices · Canada · all filters

ProductGuru Recalls republishes official safety information with attribution. We are not the issuing authority and this page is not advice. If this product is a medical device or medication, speak to a doctor or pharmacist. Found an error? Report it — corrections are published.