ProductGuru Recalls

Official recall record

BP-200 Plus Blood Pressure Monitor

Health Canada — Medical Devices · 2024-05-23 · CanadaType II
Issuing authority
Health Canada — Medical Devices
Published
2024-05-23
Country of alert
Canada
Alert level
Type II
Alert type
recall

What the authority reported

Masimo identified TC- I reusable tip clip sensors with higher-than-expected variability in SpM2 readings that are outside of the accuracy specification. Masimo's investigation found the TC-I sensors had higher variability with an overall arms of 4.4% compared to the specification of 3.5% arms. These sensors are included with the BP-200 plus device with the SpO2 option. Recall start date: March 14, 2024

What to do

Follow the instructions in the official notice — recall remedies (refund, repair, replacement, or disposal) are defined by the issuing authority and the responsible business, and they differ per recall. Do not rely on summaries, including this one.

Verify at the source: Health Canada — Medical Devices — official record →
Record last checked against the registry on 2026-05-31.

More recalls: Health Canada — Medical Devices · Canada · all filters