Official recall record
BP-200 Plus Blood Pressure Monitor
- Issuing authority
- Health Canada — Medical Devices
- Published
- 2024-05-23
- Country of alert
- Canada
- Alert level
- Type II
- Alert type
- recall
What the authority reported
Masimo identified TC- I reusable tip clip sensors with higher-than-expected variability in SpM2 readings that are outside of the accuracy specification. Masimo's investigation found the TC-I sensors had higher variability with an overall arms of 4.4% compared to the specification of 3.5% arms. These sensors are included with the BP-200 plus device with the SpO2 option. Recall start date: March 14, 2024
What to do
Follow the instructions in the official notice — recall remedies (refund, repair, replacement, or disposal) are defined by the issuing authority and the responsible business, and they differ per recall. Do not rely on summaries, including this one.
Verify at the source:
Health Canada — Medical Devices — official record →
Record last checked against the registry on 2026-05-31.
Record last checked against the registry on 2026-05-31.
ProductGuru Recalls republishes official safety information with attribution. We are not
the issuing authority and this page is not advice. If this product is a medical device or
medication, speak to a doctor or pharmacist. Found an error?
Report it — corrections are published.