Official recall record
BD Multitest™ 6-color TBNK (CE/IVD)
- Issuing authority
- Health Canada — Medical Devices
- Published
- 2024-05-23
- Country of alert
- Canada
- Alert level
- Type II
- Alert type
- recall
What the authority reported
BD observed an increasing trend of complaints for unusual staining pattern, nonspecific aggregates, high background, double positive signal, and/or antibodies contaminated with substance for the affected catalog. The investigation detected a damaged component in affected product catalog may affect integrity of stained cells, causing increase in cell debris (i.e. background) as shown on SSC VS CD45 positive leukocyte flow cytometry plots, resulting in atypical increase of double positive events on CD3+CD4+ vs CD3+CD8+ and/or CD3-CD16+CD56+ vs CD3-CD19+ plots. Recall start date: May 8, 2024
What to do
Follow the instructions in the official notice — recall remedies (refund, repair, replacement, or disposal) are defined by the issuing authority and the responsible business, and they differ per recall. Do not rely on summaries, including this one.