Official recall record
Fabian™ HFO Ventilator
- Issuing authority
- Health Canada — Medical Devices
- Published
- 2024-05-24
- Country of alert
- Canada
- Alert level
- Type I
- Alert type
- recall
What the authority reported
Software V5.2.2 changed the criteria for detecting a patient circuit disconnect during HFO ventilation with the intent to reduce the potential for patient safety-related risks. While ventilating in HFO mode with pressure amplitude (Pamp) > 1.5x mean pressure (Pmean) and with the breathing circuit properly connected, if the breath-by-breath leak monitor exceeds 40%, the ventilator triggers the disconnect alarm. the displayed volume and leak monitors reset to zero and the ventilator reduces the HFOV amplitude as a protection mechanism to prevent lung injury upon reconnection following a true disconnection. If the patient is not disconnected, this results in a small increase in Pmean, and potentially large reduction in amplitude until the alarm is cleared. The potential health risk for a false disconnect alarm occurring that cannot be resolved is inadequate ventilation due to the oscillation being out of range, resulting in hypoxia, hypercapnia, hypoventilation and/or respiratory arrest. Recall start date: April 17, 2024
What to do