Noblus Diagnostic Ultrasound Scanner - Main Unit

Health Canada — Medical Devices · 2024-06-05 · CanadaType II
Issuing authority
Health Canada — Medical Devices
Published
2024-06-05
Country of alert
Canada
Alert level
Type II
Alert type
recall

What the authority reported

Fujifilm has become aware that some of the AC adapters of Noblus heat up and cause smoke, in the worst case, when excessive mechanical stress is applied to the socket for power cable (frequent insertion/removal of the power cable). This mechanical stress would cause failure of the solder joint inside of the AC adapter and generates excessive heat. Recall Start Date: May 9, 2024