Howmedica Revision Instrument System (HRIS) Storage & Sterilization Case and Tray

Health Canada — Medical Devices · 2024-06-06 · CanadaType II
Issuing authority
Health Canada — Medical Devices
Published
2024-06-06
Country of alert
Canada
Alert level
Type II
Alert type
recall

What the authority reported

Gray revision instruments are single-use, sterile-packaged manual surgical instruments intended for one-time use during hip or knee arthroplasty, or trauma surgeries. The HRIS storage & sterilization case and tray part number 6210-9-900 has dedicated locations that include instrument silhouettes and part number markings that are silkscreen printed on the trays for the instruments. This may lead to the re-use and re-sterilization of these single-use sterile-packaged instruments. Recall Start Date: May 23, 2024