ProductGuru Recalls

Official recall record

AXIOM Vertix and AXIOM Multix

Health Canada — Medical Devices · 2024-06-14 · CanadaType II
Issuing authority
Health Canada — Medical Devices
Published
2024-06-14
Country of alert
Canada
Alert level
Type II
Alert type
recall

What the authority reported

Siemens Healthineers was informed by the monitor supplier that 5 out of 27.000 delivered power supplies of a touch-display power supply unit were affected by overheating due to a short circuit. In one of the five events, the connected monitor caught fire. No injuries were associated with these events. To ensure the continued safe operation of your system and to safeguard its lifetime, the power supply of the touch display will be replaced preventively. Recall start date: May 29, 2024

What to do

Follow the instructions in the official notice — recall remedies (refund, repair, replacement, or disposal) are defined by the issuing authority and the responsible business, and they differ per recall. Do not rely on summaries, including this one.

Verify at the source: Health Canada — Medical Devices — official record →
Record last checked against the registry on 2026-05-31.

More recalls: Health Canada — Medical Devices · Canada · all filters

ProductGuru Recalls republishes official safety information with attribution. We are not the issuing authority and this page is not advice. If this product is a medical device or medication, speak to a doctor or pharmacist. Found an error? Report it — corrections are published.