Flowpra Class I And II Screening Tests

Health Canada — Medical Devices · 2024-06-20 · CanadaType II
Issuing authority
Health Canada — Medical Devices
Published
2024-06-20
Country of alert
Canada
Alert level
Type II
Alert type
recall

What the authority reported

A suspected false negative for HLA class I with FL1-30 lot 022. Complaint investigation testing was performed using 4 samples with low levels of A2 antibody specificities and 1 sample with low levels of A31 antibody specificities. Investigation confirmed a lower percentage of positive reactions for samples tested with FL1-30 lot 022. In addition, investigation noted that lot specific user document, TDX-OLI-DMR-LOT-8536 revision 01, has incorrect reference values for the distribution of certain HLA class I antigen specificities. Recall start date: May 29, 2024