Official recall record
Easy Moving Plus B Mobile X-Ray Unit With Digital Detector
- Issuing authority
- Health Canada — Medical Devices
- Published
- 2024-06-20
- Country of alert
- Canada
- Alert level
- Type II
- Alert type
- recall
What the authority reported
Notification to inform customers regarding potential safety issue related to the steel cable installed inside the mobile column which supports the weight of telescopic arm with the tube head. The potential risk consists of an interference which could cause that the safety system which blocks the arm of the cable is cut off does not work properly. The interference occurs between the cable and a metal plate of the arm's carriage due to shape of the cable's crimped terminals. Scope of the safety recall is to review the installed base according to the manufacturer's instructions and install the kit immediately if the cable could interfere with the parachute system. Recall start date: June 5, 2024
What to do
Follow the instructions in the official notice — recall remedies (refund, repair, replacement, or disposal) are defined by the issuing authority and the responsible business, and they differ per recall. Do not rely on summaries, including this one.