Horizon X-Ray Bone Densitometer (Dxa)
- Issuing authority
- Health Canada — Medical Devices
- Published
- 2024-06-21
- Country of alert
- Canada
- Alert level
- Type I
- Alert type
- recall
What the authority reported
During standard safety design verification tests, Hologic has identified a non-conformance in Horizon DXA devices. The non-conformance pertains to electromagnetic compatibility requirements according to the international technical standard IEC 60601 - 1 - 2 for the safety and essential performance of medical electrical equipment, where the result from the Horizon DXA system exceeded the limit for acceptability. Recall start date: June 14, 2024
What to do
Follow the instructions in the official notice — recall remedies (refund, repair, replacement, or disposal) are defined by the issuing authority and the responsible business, and they differ per recall. Do not rely on summaries, including this one.