Infinity 5 and 7 Implantable Pulse Generator
- Issuing authority
- Health Canada — Medical Devices
- Published
- 2024-07-03
- Country of alert
- Canada
- Alert level
- Type II
- Alert type
- recall
What the authority reported
Abbott has initiated a voluntary medical device recall of Implantable Pulse Generator (IPG) for patients with non-rechargeable Infinity™ Deep Brain Stimulation (DBS) neurostimulation systems, as the duration between the IPG reaching battery elective replacement indicator (ERI) threshold and end of service (EOS) may be 45-55% shorter than indicated in the product labeling. Recall Start Date: June 6, 2024