Official recall record
Proclaim Implantable Pulse Generators
- Issuing authority
- Health Canada — Medical Devices
- Published
- 2024-07-04
- Country of alert
- Canada
- Alert level
- Type II
- Alert type
- recall
What the authority reported
Abbott has initiated a voluntary medical device recall of Implantable Pulse Generator (IPG) for patients with non-rechargeable proclaim neurostimulation systems as the duration between the IPG reaching battery elective replacement indicator (ERI) threshold and end of service (EOS) may be 45-55% shorter than indicated in the product labeling. Recall Start Date: June 6, 2024
What to do
Follow the instructions in the official notice — recall remedies (refund, repair, replacement, or disposal) are defined by the issuing authority and the responsible business, and they differ per recall. Do not rely on summaries, including this one.
Verify at the source:
Health Canada — Medical Devices — official record →
Record last checked against the registry on 2026-05-31.
Record last checked against the registry on 2026-05-31.
ProductGuru Recalls republishes official safety information with attribution. We are not
the issuing authority and this page is not advice. If this product is a medical device or
medication, speak to a doctor or pharmacist. Found an error?
Report it — corrections are published.