ProductGuru Recalls

Official recall record

CALiaGold PierceTube

Health Canada — Medical Devices · 2024-07-04 · CanadaType II
Issuing authority
Health Canada — Medical Devices
Published
2024-07-04
Country of alert
Canada
Alert level
Type II
Alert type
recall

What the authority reported

Following a customer complaint, Sentinel detected a defect in three (3) lots of the collection tube CALiaGold PierceTube, ref. 1151300: lot 31390, 40037, 40045. The defect is an inner plastic crack that connects the two chambers of the collection tube, causing a reduction of the buffer volume needed for the measurement. It cannot be excluded that a significant volume decrease may lead to the impossibility to perform the test. The root cause of the plastic crack is still under investigation. Recall Start Date: June 13, 2024

What to do

Follow the instructions in the official notice — recall remedies (refund, repair, replacement, or disposal) are defined by the issuing authority and the responsible business, and they differ per recall. Do not rely on summaries, including this one.

Verify at the source: Health Canada — Medical Devices — official record →
Record last checked against the registry on 2026-05-31.

More recalls: Health Canada — Medical Devices · Canada · all filters

ProductGuru Recalls republishes official safety information with attribution. We are not the issuing authority and this page is not advice. If this product is a medical device or medication, speak to a doctor or pharmacist. Found an error? Report it — corrections are published.