Official recall record
CALiaGold PierceTube
- Issuing authority
- Health Canada — Medical Devices
- Published
- 2024-07-04
- Country of alert
- Canada
- Alert level
- Type II
- Alert type
- recall
What the authority reported
Following a customer complaint, Sentinel detected a defect in three (3) lots of the collection tube CALiaGold PierceTube, ref. 1151300: lot 31390, 40037, 40045. The defect is an inner plastic crack that connects the two chambers of the collection tube, causing a reduction of the buffer volume needed for the measurement. It cannot be excluded that a significant volume decrease may lead to the impossibility to perform the test. The root cause of the plastic crack is still under investigation. Recall Start Date: June 13, 2024