Alinity HQ Analyzer (ref 75791)

Health Canada — Medical Devices · 2024-07-05 · CanadaType II
Issuing authority
Health Canada — Medical Devices
Published
2024-07-05
Country of alert
Canada
Alert level
Type II
Alert type
recall

What the authority reported

Abbott has identified the following issues when using Alinity HQ Analyzer software version 5.6 and below which will be corrected in software version 5.9. Issue 1: When processing a CBC+diff sample that immediately follows a CBC+diff+retic sample, diluent may backflow into the RBC dilution cup. This may result in falsely low red blood cell (RBC) results and consequently falsely high mean cell hemoglobin concentration (MCHC), falsely high mean cell hemoglobin (MCH), and falsely low hematocrit (HCT). Issue 2: An overestimation of the basophil (baso) count may occur on some samples when cell events are incorrectly counted as baso during sample analysis. this may result in falsely increased baso counts and %b (%baso). Recall start date: June 21, 2024

What to do

Follow the instructions in the official notice — recall remedies (refund, repair, replacement, or disposal) are defined by the issuing authority and the responsible business, and they differ per recall. Do not rely on summaries, including this one.

Verify at the source: Health Canada — Medical Devices — official record →
Record last checked against the registry on 2026-05-31.

More recalls: Health Canada — Medical Devices · Canada · all filters

ProductGuru Recalls republishes official safety information with attribution. We are not the issuing authority and this page is not advice. If this product is a medical device or medication, speak to a doctor or pharmacist. Found an error? Report it — corrections are published.