Official recall record
Soltive Laser System
- Issuing authority
- Health Canada — Medical Devices
- Published
- 2024-07-12
- Country of alert
- Canada
- Alert level
- Type II
- Alert type
- recall
What the authority reported
Olympus investigated complaints received where customers reported difficulties in pairing the wireless footswitch with the Soltive laser, resulting in delays in surgical procedures or prolonged surgery. In an effort to mitigate any immediate and potential risk to patient health, Olympus is requiring Soltive laser users to have an Olympus wired footswitch (part number TFL-AFSW) available for immediate use as a backup in the event a wireless footswitch does not pair with the Soltive laser. Recall start date: June 28, 2024