BIOFIRE® Filmarray® Gastrointestinal (GI) Panel

Health Canada — Medical Devices · 2024-07-18 · CanadaType II
Issuing authority
Health Canada — Medical Devices
Published
2024-07-18
Country of alert
Canada
Alert level
Type II
Alert type
recall

What the authority reported

An increased number of customer complaints associated with Vibrio/Vibrio Cholerae false positive detections by the BIOFIRE Filmarray Gastrointestinal (GI) Panel have been observed when using Remel Cary-Blair transport medium. The BIOFIRE GI panel and the Cary-Blair transport medium are performing as intended. Per INV-19261, the root cause of the false positive Vibrio/Vibrio Cholerae results detected by the BIOFIRE GI panel is the presence of non-viable organism in the Remel Cary-Blair transport medium. Recall Start Date: July 4, 2024