Trilogy EVO and EV300 (ref 75855)

Health Canada — Medical Devices · 2024-07-24 · CanadaType II
Issuing authority
Health Canada — Medical Devices
Published
2024-07-24
Country of alert
Canada
Alert level
Type II
Alert type
recall

What the authority reported

Philips Respironics has become aware that Trilogy EVO and Trilogy EV300 devices do not comply with the obstruction alarm requirements specified within ISO 80601-2-12 clause 201.12.4.108 and ISO 80601-2-72 clause 201.12.4.107. The standards specify that the maximum delay from obstruction to alarm shall be no more than 2 breath cycles or 5 seconds, whichever is greater. The current obstruction alarm delay for these devices has been measured up to 65 seconds in certain bench test conditions. This is a delay potentially 60 seconds greater than required by the standards. Recall start date: July 1, 2024