ProductGuru Recalls

Official recall record

VENTANA® anti-Cd10 (SP67) Rabbit Monoclonal Primary Antibody

Health Canada — Medical Devices · 2024-07-25 · CanadaType II
Issuing authority
Health Canada — Medical Devices
Published
2024-07-25
Country of alert
Canada
Alert level
Type II
Alert type
recall

What the authority reported

Roche has received complaints regarding high background and off-target staining with certain lots of the VENTANA® anti-CD10 (SP67) Rabbit Monoclonal Primary Antibody resulting in a risk of false positive results. Subsequent investigation identified that this high background and off-target staining can interfere with slide interpretation when using the recommended method sheet OptiView DAB IHC Detection Kit (product no.760-700 / 06396500001) / OptiView Amplification Kit (product no. 760-099 / 06396518001) and the ultraView Universal DAB Detection Kit (product no.760-500 i 05269806001) /Amplification Kit (product no. 760-080 / 05266114001) protocol. Recall start date: July 11, 2024

What to do

Follow the instructions in the official notice — recall remedies (refund, repair, replacement, or disposal) are defined by the issuing authority and the responsible business, and they differ per recall. Do not rely on summaries, including this one.

Verify at the source: Health Canada — Medical Devices — official record →
Record last checked against the registry on 2026-05-31.

More recalls: Health Canada — Medical Devices · Canada · all filters

ProductGuru Recalls republishes official safety information with attribution. We are not the issuing authority and this page is not advice. If this product is a medical device or medication, speak to a doctor or pharmacist. Found an error? Report it — corrections are published.