ProductGuru Recalls

Official recall record

IRISpec CA/CB/CC (ref 75879)

Health Canada — Medical Devices · 2024-07-26 · CanadaType II
Issuing authority
Health Canada — Medical Devices
Published
2024-07-26
Country of alert
Canada
Alert level
Type II
Alert type
recall

What the authority reported

Beckman Coulter has become aware of an increase in customer complaints for intermittent IRISpec CB glucose control failure, resulting in false positive results when used in conjunction with iChemVELOCITY Urine Chemistry Strips PNS 800-7204 and 800- 7212. If a QC run fails, the system will not allow the operator to run patient samples. The QC run should be repeated and pass successfully before running patient samples. Recall start date: July 16, 2024

What to do

Follow the instructions in the official notice — recall remedies (refund, repair, replacement, or disposal) are defined by the issuing authority and the responsible business, and they differ per recall. Do not rely on summaries, including this one.

Verify at the source: Health Canada — Medical Devices — official record →
Record last checked against the registry on 2026-05-31.