Official recall record
Accu-Chek Guide
- Issuing authority
- Health Canada — Medical Devices
- Published
- 2024-07-31
- Last updated (per notice)
- 2024-09-26
- Country of alert
- Canada
- Alert level
- Type II
- Alert type
- recall
What the authority reported
The issue arises when a blood glucose measurement is taken and the result is displayed in the wrong measuring unit (i.e., mg/dl instead of mmol/l). If it goes unnoticed, the incorrectly displayed blood glucose value may result in inappropriate therapy decisions, including the administration of insulin, and could potentially result in severe consequences. Such possible consequences range from no clinical impact to adverse health events such as hypoglycemia. Recall start date: May 14, 2024