ProductGuru Recalls

Official recall record

Power Express Centrifuge

Health Canada — Medical Devices · 2024-08-01 · CanadaType II
Issuing authority
Health Canada — Medical Devices
Published
2024-08-01
Country of alert
Canada
Alert level
Type II
Alert type
recall

What the authority reported

Customers have observed an issue where the metal part of the sample holder detaches from the plastic holder base when the robot arm grips a sample tube. If this occurs at a centrifuge, the metal part can inadvertently enter the centrifuge along with the sample tube. The centrifuge is unable to detect the presence of the detached metal part and begins to rotate. This can pose risks to both the equipment and the samples being processed. Recall start date: July 22, 2024

What to do

Follow the instructions in the official notice — recall remedies (refund, repair, replacement, or disposal) are defined by the issuing authority and the responsible business, and they differ per recall. Do not rely on summaries, including this one.

Verify at the source: Health Canada — Medical Devices — official record →
Record last checked against the registry on 2026-05-31.

More recalls: Health Canada — Medical Devices · Canada · all filters

ProductGuru Recalls republishes official safety information with attribution. We are not the issuing authority and this page is not advice. If this product is a medical device or medication, speak to a doctor or pharmacist. Found an error? Report it — corrections are published.