ProductGuru Recalls

Official recall record

Karl Storz Telescopes

Health Canada — Medical Devices · 2024-08-01 · CanadaType II
Issuing authority
Health Canada — Medical Devices
Published
2024-08-01
Last updated (per notice)
2024-08-27
Country of alert
Canada
Alert level
Type II
Alert type
recall

What the authority reported

The IFUs for the devices identifiers previously included cleaning methods other than manual cleaning and pre-vacuum steam as the approved reprocessing modalities. The IFU was updated to only include these two methods moving forward; therefore, there is a risk that a reprocessing modality that is not currently cleared could be utilized for reprocessing these devices. Recall start date: July 25, 2024

What to do

Follow the instructions in the official notice — recall remedies (refund, repair, replacement, or disposal) are defined by the issuing authority and the responsible business, and they differ per recall. Do not rely on summaries, including this one.

Verify at the source: Health Canada — Medical Devices — official record →
Record last checked against the registry on 2026-05-31.

More recalls: Health Canada — Medical Devices · Canada · all filters