Blue Line Uncuffed Tracheostomy Tube - Single Use

Health Canada — Medical Devices · 2024-08-02 · CanadaType I
Issuing authority
Health Canada — Medical Devices
Published
2024-08-02
Country of alert
Canada
Alert level
Type I
Alert type
recall

What the authority reported

Manufacturer has identified an issue related to the neck plate/flange of the device. Specifically, this failure mode can manifest itself during use as a complete or partial detachment of the neck plate from the tracheostomy tube on the device. This failure mode can lead to inadequate ventilation for the patient and complete dislodgement of the tracheostomy tube. Hypoxia, underdose, cardiopulmonary collapse, bradycardia, hypotension, respiratory arrest, or asphyxia can potentially result from the partial or complete detachment of the flange. To date, manufacturer has received five (5) reports of serious injury, and zero (0) deaths potentially related to this issue. Recall start date: July 29, 2024

What to do

Follow the instructions in the official notice — recall remedies (refund, repair, replacement, or disposal) are defined by the issuing authority and the responsible business, and they differ per recall. Do not rely on summaries, including this one.

Verify at the source: Health Canada — Medical Devices — official record →
Record last checked against the registry on 2026-05-31.

More recalls: Health Canada — Medical Devices · Canada · all filters

ProductGuru Recalls republishes official safety information with attribution. We are not the issuing authority and this page is not advice. If this product is a medical device or medication, speak to a doctor or pharmacist. Found an error? Report it — corrections are published.