Official recall record
StealthStation S8 Application
- Issuing authority
- Health Canada — Medical Devices
- Published
- 2024-08-14
- Country of alert
- Canada
- Alert level
- Type II
- Alert type
- recall
What the authority reported
Medtronic is writing to customers to inform them of an issue associated with the StealthStation™ S8 application, and specifically the display that shows the "distance to target" or "tip stop point" measurements. There is the potential for a missing digit, letter, or decimal in either the "distance to target" or "tip stop point" display during navigation. In certain situations, this anomaly could lead to incorrect data being displayed during surgical procedures. Recall Start Date: April 10, 2024
What to do
Follow the instructions in the official notice — recall remedies (refund, repair, replacement, or disposal) are defined by the issuing authority and the responsible business, and they differ per recall. Do not rely on summaries, including this one.