Wavewriter Alpha Implantable Pulse Generator Kit

Health Canada — Medical Devices · 2024-08-14 · CanadaType II
Issuing authority
Health Canada — Medical Devices
Published
2024-08-14
Country of alert
Canada
Alert level
Type II
Alert type
recall

What the authority reported

Boston Scientific has received reports of undesired sensation and/or awareness of IPG turning off and on during IPG charging for patients with rechargeable SCS IPGs. Investigation of these reports determined that the device may reset (by design) due to detection of noise or interference during IPG charging. When the system resets, the patient's programmed stimulation turns off for approximately 10-15 seconds and then turns back on. When this occurs, some patients may perceive the sudden changes in stimulation. As soon as the reset is complete, the IPG resumes normal operation, including delivery of stimulation therapy. Boston Scientific has identified improvements (including an IPG firmware update for implanted devices) that will prevent this device reset behavior from occurring during IPG charging. Updating the IPG firmware eliminates the possibility of a coincident routine system check during IPG charging, thus preventing a potential system reset. This firmware update is available for any rechargeable SCS IPGs identified. Recall Start Date: July 17, 2024

What to do

Follow the instructions in the official notice — recall remedies (refund, repair, replacement, or disposal) are defined by the issuing authority and the responsible business, and they differ per recall. Do not rely on summaries, including this one.

Verify at the source: Health Canada — Medical Devices — official record →
Record last checked against the registry on 2026-05-31.

More recalls: Health Canada — Medical Devices · Canada · all filters

ProductGuru Recalls republishes official safety information with attribution. We are not the issuing authority and this page is not advice. If this product is a medical device or medication, speak to a doctor or pharmacist. Found an error? Report it — corrections are published.