Fetalscreen II

Health Canada — Medical Devices · 2024-08-15 · CanadaType II
Issuing authority
Health Canada — Medical Devices
Published
2024-08-15
Country of alert
Canada
Alert level
Type II
Alert type
recall

What the authority reported

A customer complaint investigation has confirmed that the lot of Fetalscreen II stated above shows unexpected macroscopic reactivity (agglutination) when performing QC testing with the positive control cells. This will impact the reading of rosettes to validate the positive control and may impact the reading of positive tests, potentially resulting in false negative results. Recall Start Date: July 22, 2024

What to do

Follow the instructions in the official notice — recall remedies (refund, repair, replacement, or disposal) are defined by the issuing authority and the responsible business, and they differ per recall. Do not rely on summaries, including this one.

Verify at the source: Health Canada — Medical Devices — official record →