Guidewire 70cm X 1mm (0.038 Inch)

Health Canada — Medical Devices · 2024-08-16 · CanadaType II
Issuing authority
Health Canada — Medical Devices
Published
2024-08-16
Country of alert
Canada
Alert level
Type II
Alert type
recall

What the authority reported

BD has received multiple reports that guidewire code 5583705, which is intended for use with the Bard 14.5 Fr Long-Term Dual Lumen Hemodialysis Catheters, may have been misassembled within the hoop such that the proximal (stiff) end of the guidewire is located at the distal end of the hoop. When loaded correctly the distal (flexible) end of the guidewire should be located at the distal end of the guidewire hoop. Recall Start Date: August 1, 2024