Official recall record
CARDIOVIT® AT-180 Electrocardiographs
- Issuing authority
- Health Canada — Medical Devices
- Published
- 2024-08-20
- Country of alert
- Canada
- Alert level
- Type II
- Alert type
- recall
What the authority reported
Occasionally high-frequency signal artifacts are recorded during an ECG acquisition performed by CARDIOVIT® AT-180 electrocardiographs. It has been observed that these high-frequency artifacts have been occasionally incorrectly identified as pacemaker spikes by the electrocardiograph. Recall start date: July 24, 2024