Liaison® Testosterone XT
- Issuing authority
- Health Canada — Medical Devices
- Published
- 2024-08-23
- Country of alert
- Canada
- Alert level
- Type II
- Alert type
- recall
What the authority reported
The affected kit lots of Liaison® Testosterone XT may produce falsely low results in samples below 2 ng/ml. Median values below 2 ng/ml show a negative bias to the median values listed in the IFU. The negative bias in samples below 2 ng/ml may impact interpretation of results relative to the reference range claims for the adult female populations and pediatric populations. Recall start date: July 29. 2024