Official recall record
Quadrox-I Adult Oxygenator W/Integrated Hardshell Reservoir
- Issuing authority
- Health Canada — Medical Devices
- Published
- 2024-08-30
- Country of alert
- Canada
- Alert level
- Type II
- Alert type
- recall
What the authority reported
Maquet Cardiopulmonary's monthly inspection of oxygenators for the presence of the Softline coating revealed that two out of ten oxygenators tested did not pass. The Softline coating is a vital component in the biocompatibility of blood contact components used in extracorporeal circulation. This means that the oxygenators installed in the products listed above do not have the Softline coating. The issue is limited to one lot number in Canada. Recall Start Date: August 20, 2024
What to do
Follow the instructions in the official notice — recall remedies (refund, repair, replacement, or disposal) are defined by the issuing authority and the responsible business, and they differ per recall. Do not rely on summaries, including this one.
Record last checked against the registry on 2026-05-31.