ProductGuru Recalls

Official recall record

Leica Cryostats

Health Canada — Medical Devices · 2024-09-06 · CanadaType II
Issuing authority
Health Canada — Medical Devices
Published
2024-09-06
Country of alert
Canada
Alert level
Type II
Alert type
recall

What the authority reported

On August 6, 2024, Leica Biosystems became aware, through post market surveillance, of a customer using flammable freezing spray within the cryostat chamber. The air/spray mixture in the chamber ignited, leading to an explosion causing slight burns to the arms and face of the technician. The scope of the recall has been expanded to include all devices produced prior to 2009/01/01. Recall start date: August 29, 2024

What to do

Follow the instructions in the official notice — recall remedies (refund, repair, replacement, or disposal) are defined by the issuing authority and the responsible business, and they differ per recall. Do not rely on summaries, including this one.

Verify at the source: Health Canada — Medical Devices — official record →
Record last checked against the registry on 2026-05-31.

More recalls: Health Canada — Medical Devices · Canada · all filters