30 atm MONARCH® Inflation Syringe

Health Canada — Medical Devices · 2024-09-09 · CanadaType II
Issuing authority
Health Canada — Medical Devices
Published
2024-09-09
Country of alert
Canada
Alert level
Type II
Alert type
recall

What the authority reported

On July 31, 2024, Merit internally identified an issue related to potential holes being found in the Tyvek® sterile barrier of certain MONARCH® inflation devices during the manufacture of lot H3003824. Visual inspection of the packaged units from lot H3003824 found a potential hole in 28 of 776 manufactured units. The 28 units underwent a bubble test where 23 of the 28 units failed, confirming a hole in the sterile barrier. Further investigation has determined that certain placements of the MONARCH® inflation device within a rigid formed blister tray can cause a hole in the Tyvek®. Should a MONARCH® inflation device's placement result in the unit's retainer cap being positioned above the rigid blister flange, the Tyvek® can contact a multivac's cross-cut station potentially leading to an abrasion or hole in the Tyvek®. Recall start date: August 28, 2024

What to do

Follow the instructions in the official notice — recall remedies (refund, repair, replacement, or disposal) are defined by the issuing authority and the responsible business, and they differ per recall. Do not rely on summaries, including this one.

Verify at the source: Health Canada — Medical Devices — official record →
Record last checked against the registry on 2026-05-31.

More recalls: Health Canada — Medical Devices · Canada · all filters

ProductGuru Recalls republishes official safety information with attribution. We are not the issuing authority and this page is not advice. If this product is a medical device or medication, speak to a doctor or pharmacist. Found an error? Report it — corrections are published.