IntelliVue MX550 Patient Monitor

Health Canada — Medical Devices · 2024-09-18 · CanadaType II
Issuing authority
Health Canada — Medical Devices
Published
2024-09-18
Country of alert
Canada
Alert level
Type II
Alert type
recall

What the authority reported

Philips became aware of one IntelliVue power supply with a broken ground bolt upon disconnection of the ground cable from the equipotential ground connector. Loss of electrical grounding may negatively affect the device's electromagnetic immunity and emission. Degraded electromagnetic immunity can cause the monitor to generate unreadable or unusable waveforms, potentially leading to incorrect/delayed patient treatment. Excessive or unintended electromagnetic emissions can also negatively impact the function of equipment in the vicinity of the patient monitor, potentially leading to delayed procedure. Although unlikely, these scenarios could potentially result in patient harm. Recall Start Date: September 4, 2024

What to do

Follow the instructions in the official notice — recall remedies (refund, repair, replacement, or disposal) are defined by the issuing authority and the responsible business, and they differ per recall. Do not rely on summaries, including this one.

Verify at the source: Health Canada — Medical Devices — official record →
Record last checked against the registry on 2026-05-31.