ProductGuru Recalls

Official recall record

MEERA Operating Tables

Health Canada — Medical Devices · 2024-09-19 · CanadaType II
Issuing authority
Health Canada — Medical Devices
Published
2024-09-19
Country of alert
Canada
Alert level
Type II
Alert type
recall

What the authority reported

It is possible that the error code "50037" occurs on the IR-Hand Control when a MEERA-table is controlled, and the table stops the movement. This error may occur sporadically when the MEERA-table is controlled via an IR-Hand Control. With a cable-hand control the error does not occur. If the error "50037" occurs the movement is not permanently blocked. Recall Start Date: September 11, 2024

What to do

Follow the instructions in the official notice — recall remedies (refund, repair, replacement, or disposal) are defined by the issuing authority and the responsible business, and they differ per recall. Do not rely on summaries, including this one.

Verify at the source: Health Canada — Medical Devices — official record →
Record last checked against the registry on 2026-05-31.

More recalls: Health Canada — Medical Devices · Canada · all filters

ProductGuru Recalls republishes official safety information with attribution. We are not the issuing authority and this page is not advice. If this product is a medical device or medication, speak to a doctor or pharmacist. Found an error? Report it — corrections are published.