BD Synapsys™ Informatics Solution
- Issuing authority
- Health Canada — Medical Devices
- Published
- 2024-09-26
- Country of alert
- Canada
- Alert level
- Type II
- Alert type
- recall
What the authority reported
BD identified through one (1) complaint that for BD Kiestra™ customers with BD Synapsys™, running Synapsys™ software versions 4.20, 4.30, 5.10, 5.20, or 5.30 and utilizing the in-line breakpoint feature, an error is occurring where BD Synapsys™ will incorrectly display breakpoints for organisms that do not have defined breakpoints in CLSI or EUCAST. This occurs when the identified organism is included in the same interpretive group as a species for which there is a specific breakpoint. Recall start date: September 12, 2024
What to do
Follow the instructions in the official notice — recall remedies (refund, repair, replacement, or disposal) are defined by the issuing authority and the responsible business, and they differ per recall. Do not rely on summaries, including this one.