ProductGuru Recalls

Official recall record

Trilogy Evo and Trilogy Ev300

Health Canada — Medical Devices · 2024-10-02 · CanadaType I
Issuing authority
Health Canada — Medical Devices
Published
2024-10-02
Country of alert
Canada
Alert level
Type I
Alert type
recall

What the authority reported

Philips Respironics has become aware of a potential issue when using in-line nebulizers in certain configurations with Trilogy Evo and Trilogy Ev300 devices. The use of in-line nebulizers placed in certain locations can result in aerosol deposits accumulating over time on the device's internal flow sensor. Impacted flow sensors may result in inaccurate flow measurements in certain circumstances. Recall start date: September 18, 2024

What to do

Follow the instructions in the official notice — recall remedies (refund, repair, replacement, or disposal) are defined by the issuing authority and the responsible business, and they differ per recall. Do not rely on summaries, including this one.

Verify at the source: Health Canada — Medical Devices — official record →
Record last checked against the registry on 2026-05-31.

More recalls: Health Canada — Medical Devices · Canada · all filters