ProductGuru Recalls

Official recall record

Sterilizable Internal Defibrillation Paddles

Health Canada — Medical Devices · 2024-10-03 · CanadaType II
Issuing authority
Health Canada — Medical Devices
Published
2024-10-03
Country of alert
Canada
Alert level
Type II
Alert type
recall

What the authority reported

Stryker has received complaints indicating issues with the internal paddles. Complaints which state that the paddles are not reaching the expected cleaning and sterilization cycle count designated in the instructions for use and are showing signs of cracking. A CAPA was opened November 2023. Recall start date: September 20, 2024

What to do

Follow the instructions in the official notice — recall remedies (refund, repair, replacement, or disposal) are defined by the issuing authority and the responsible business, and they differ per recall. Do not rely on summaries, including this one.

Verify at the source: Health Canada — Medical Devices — official record →
Record last checked against the registry on 2026-05-31.

More recalls: Health Canada — Medical Devices · Canada · all filters