Epoc Bgem Bun Test Card [25pk]

Health Canada — Medical Devices · 2024-10-04 · CanadaType II
Issuing authority
Health Canada — Medical Devices
Published
2024-10-04
Country of alert
Canada
Alert level
Type II
Alert type
recall

What the authority reported

Siemens Healthcare Diagnostics Inc. has confirmed that a negative sodium bias may exist in some card lots approximately 2 months prior to expiration. The observed maximum negative bias from internal testing is -10 mmol/l. No positive bias has been observed. Quality Control (QC) fluid may or may not detect the issue depending on the bias magnitude and the frequency of QC testing. Please note the card lots identified are now expired. Recall start date: September 25, 2024