Ultraview DM3 Monitor

Health Canada — Medical Devices · 2024-10-09 · CanadaType II
Issuing authority
Health Canada — Medical Devices
Published
2024-10-09
Country of alert
Canada
Alert level
Type II
Alert type
recall

What the authority reported

The lithium-ion battery used in the DM3 is susceptible to failure and overheating when used beyond its recommended lifetime or otherwise damaged. This is due to a short in the battery with indeterminant cause, but likely attributed to use beyond service life, overcharging, or damage to the battery. Batteries known to be older than two years or batteries that are no longer holding a charge should be replaced. Damaged batteries must be replaced immediately. Recall Start Date: September 27, 2024