Official recall record
MiniMed 670 g, 780 g, 770 g and 630 g
- Issuing authority
- Health Canada — Medical Devices
- Published
- 2024-10-16
- Last updated (per notice)
- 2024-10-18
- Country of alert
- Canada
- Alert level
- Type I
- Alert type
- recall
What the authority reported
In some instances, pumps that have been dropped, bumped, or experienced physical impact, may experience damage to internal electrical components, which may result in reduced battery life, causing the battery alerts to occur when less battery life remains than the user guide states. Recall start date: October 11, 2024
What to do
Follow the instructions in the official notice — recall remedies (refund, repair, replacement, or disposal) are defined by the issuing authority and the responsible business, and they differ per recall. Do not rely on summaries, including this one.
Verify at the source:
Health Canada — Medical Devices — official record →
Record last checked against the registry on 2026-05-31.